Factors affecting MV imager projection offset in machine performance check for Varian linear accelerator

( views:10, downloads:1 )
Author:
No author available
Journal Title:
Chinese Journal of Medical Physics
Issue:
7
DOI:
10.3969/j.issn.1005-202X.2024.07.004
Key Word:
直线加速器;机器性能检测;机载影像系统;编码器;等中心校正;linear accelerator;machine performance check;on-board imager;encoder;isocenter calibration

Abstract: Objective To explore the main factors affecting the MV imager projection offset in the machine performance check(MPC)for Varian Vital Beam linear accelerator.Methods The MV imager projection offsets in the MPC after repairing the MV imaging arm encoder of shoulder motor,locking the treatment couch,and isocenter calibration were analyzed.Results MPC results revealed that the MV imager projection offset after repairing the MV imaging arm encoder of shoulder motor was(0.310±0.001)mm,significantly less than(0.450±0.010)mm in the blank group.The difference in MV imager projection offset between the isocenter calibration group and the blank group was trivial.The MV imager projection offset after locking the treatment couch was(0.240±0.030)mm,significantly less than(0.450±0.010)mm in the blank group.When MPC was carried out after repairing the imaging arm encoder and performing isocenter calibration,there was no significant statistical difference in MV imager center offset between the locked and unlocked treatment couch.Conclusion The damage of MV imaging arm encoder of shoulder motor is the main factor causing abnormal MV imager projection offsets.Locking the treatment couch before the MV imaging center check can reduce the results,but it cannot eliminate the MV imager projection offset.

  • This article has no references!
WanfangData CO.,Ltd All Rights Reserved
About WanfangData | Contact US
Healthcare Department, Fuxing Road NO.15, Haidian District Beijing, 100038 P.R.China
Tel:+86-010-58882616 Fax:+86-010-58882615 Email:yiyao@wanfangdata.com.cn
Baidu
map